This article is cited in 18 scientific papers (total in 18 papers)
High Temperature Apparatuses and Structures
Gamma-densitometer for studies of high-temperature metal melts
B. A. Rusanov, E. S. Baglasova, P. S. Popel', V. E. Sidorov, A. A. Sabirzyanov Urals State Pedagogical University, Ekaterinburg
Abstract:
The setup for the measurement of the density of high-temperature melts by means of penetrating gamma-radiation is described. The modern device base, which makes it possible to automatize the work, is the main difference from the earlier existing gamma-densitometers. The key characteristics of the absolute measurements with the use of very exact values of the mass attenuation coefficient for the gamma-quanta beam are described and the accuracy of measurements is estimated. The results of the density measurement of the molten aluminum, cadmium, tin, and bismuth are presented; they confirm the a priori estimate of the error.
Received: 27.01.2017 Accepted: 18.04.2017
Citation:
B. A. Rusanov, E. S. Baglasova, P. S. Popel', V. E. Sidorov, A. A. Sabirzyanov, “Gamma-densitometer for studies of high-temperature metal melts”, TVT, 56:3 (2018), 455–460; High Temperature, 56:3 (2018), 439–443
Linking options:
https://www.mathnet.ru/eng/tvt10811 https://www.mathnet.ru/eng/tvt/v56/i3/p455
|
Statistics & downloads: |
Abstract page: | 339 | Full-text PDF : | 256 | References: | 26 |
|