This article is cited in 41 scientific papers (total in 41 papers)
Laser optics 2012
Development of precision laser goniometer systems
M. N. Burnashev, P. A. Pavlov, Yu. V. Filatov Saint-Petersburg Electrotechnical University
Abstract:
The concept of constructing precision laser goniometer systems, based on integrating a ring laser and an optical angle sensor with the holographic principle of angular scale recording is considered. The concept implies the application of the cross-calibration procedure, aimed at determination of systematic components of the errors of angle sensors, used in the system. The results of the presented system studies demonstrate the error of angular measurements amounting to ~0.01''. The results of implementing the proposed concept in the creation of a standard system of the plane angle unit of rigid rotation and the measuring and computing complex for automated control of digital angle transducers with high digit capacity are briefly presented.
Keywords:
ring laser, angle measurements, cross-calibration, angle transducer, optical angle sensor.
Received: 24.10.2012 Revised: 14.11.2012
Citation:
M. N. Burnashev, P. A. Pavlov, Yu. V. Filatov, “Development of precision laser goniometer systems”, Kvantovaya Elektronika, 43:2 (2013), 130–138 [Quantum Electron., 43:2 (2013), 130–138]
Linking options:
https://www.mathnet.ru/eng/qe15045 https://www.mathnet.ru/eng/qe/v43/i2/p130
|
Statistics & downloads: |
Abstract page: | 410 | Full-text PDF : | 242 | References: | 31 | First page: | 3 |
|