ÊÎÍÅ×ÍÎ-ÝËÅÌÅÍÒÍÛÉ ÀÍÀËÈÇ ÊÎÍÑÒÐÓÊÖÈÈ
The choosing of optimal cell parameters of transcatheter aortic valve prosthesis
E. A. Ovcharenkoa, K. Yu. Klyshnikova, G. V. Savrasovb, D. V. Nyshtaevc, T. V. Glushkovaa a Research Institute for Complex Issues of Cardiovascular Diseases under the Siberian Branch of the Russian Academy of Medical Sciences, 6 Sosnovyi blvd., 650002, Kemerovo, Russia
b N. Bauman State University of Technology, 14 Brigadirsky per., Moscow, 105005, Russia
c LLC "TESIS", 18 Unnatov st., Moscow, 127083, Russia
Abstract:
This paper presents the analysis of dependences between frame basic cell geometry parameters and function via finite element analysis. The simplified models of frame cell with varied strut width, thickness and quantity in a circle was researched to evaluate radial forces, maximum stress and strain, permanent residual strain and pinching load forces. The outcomes of this study might help in the development of new artificial heart valves and during the analysis of existing in-clinical TAVI prostheses.
Keywords:
ÌÊÝ, Valve prosthesis, aorta, FEA, cell.
Received: 30.05.2014
Citation:
E. A. Ovcharenko, K. Yu. Klyshnikov, G. V. Savrasov, D. V. Nyshtaev, T. V. Glushkova, “The choosing of optimal cell parameters of transcatheter aortic valve prosthesis”, Computer Research and Modeling, 6:6 (2014), 943–954
Linking options:
https://www.mathnet.ru/eng/crm371 https://www.mathnet.ru/eng/crm/v6/i6/p943
|
Statistics & downloads: |
Abstract page: | 69 | Full-text PDF : | 40 | References: | 19 |
|