This article is cited in 11 scientific papers (total in 11 papers)
OPTO-IT
Calibration of an imaging hyperspectrometer
V. V. Podlipnovab, R. V. Skidanovab a Image Processing Systems Institute of the RAS - Branch of the FSRC "Crystallography and Photonics" RAS, Samara, Russia
b Samara National Research University, Samara, Russia
Abstract:
We describe a precise calibration procedure for an Offner imaging hyperspectrometer. We analyze the influence of the calibration on the accuracy of the spectral distribution registration. We conduct a comparative analysis of the experimental data with the results of the precise measurements of spectral distributions made using a high-precision laboratory spectrograph MS7504.
Keywords:
hyperspectrometer, calibration, tuned laser, Earth remote sensing.
Received: 25.10.2017 Accepted: 01.12.2017
Citation:
V. V. Podlipnov, R. V. Skidanov, “Calibration of an imaging hyperspectrometer”, Computer Optics, 41:6 (2017), 869–874
Linking options: